- Connect the power cord to the receptacle of the
device.
- Connect the power cord to a power socket with a
ground pin.
- If necessary, connect the potential equilization
cable of your installation system to the unit
potential equalization connector (Fig. 2-6)
- Connect the control footswitch to the footswitch
connector (Fig. 2-2).
- Position the footswitch so that it is easily accessible
for your feet.
- Insert the brackets into the bracket holders (Fig. 2-
4).
- Connect the micro-motor cord to the connector
(Fig. 1-7).
- Align the I-S
URGE
L
ED
micro-motor and cord
connector electrical contacts, then screw the motor
onto the connector.
- Connect the N
EWTRON
L
ED
scaler cord or the P
IEZOTOME
2 L
ED
handpiece cord to the connector (Fig. 1-9)
(according to option).
- Connect the N
EWTRON
L
ED
handpiece to the scaler
cord (delivered according to option).
- Check that the device is close enough to the work
area to be used without pulling on the cords. If it is
not, move the device closer.
- Place the I-S
URGE
L
ED
micro-motor on its rest.
- Place the N
EWTRON
L
ED
or P
IEZOTOME
2 L
ED
handpiece
on the handpiece support.
- Suspend the bottles or bags of physiological saline
or sterile water from the brackets.
- Adjust the position of your device to suit your angle
of view.
4. 4 FIRST USE
Before using the I
MPLANT
C
ENTER
2 for the first time, all
the equipment must be given routine maintenance
and/or sterilized according to the procedures defined
in chapter 11.
V - IRRIGATION SOLUTION
I
MPLANT
C
ENTER
2 is not designed to administer drugs. It
must be used only with bottles or bags of
physiological saline or sterile water with an individual
capacity not exceeding one litre.
VI - IRRIGATION LINES
I
MPLANT
C
ENTER
2 must be used exclusively with
SATELEC irrigation lines.
VII - ADJUSTMENTS /
SETTINGS / MODES /
INTERFACE
7. 1 STARTUP
Important:
After stopping the device, wait four seconds before
restarting it.
Each time the device is switched on, the screen
displays program P1 of the I-S
URGE
function after
displaying the welcome page.
7. 2 SETTINGS
The settings are saved when each program is
configured and are retrieved whenever that program
is selected.
7. 3 IRRIGATION FUNCTION
a) Flush / prime
The flush / prime function is available on the LCD
screen and on the footswitch.
11
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...