e) Control footswitch
Having access to a large number of functions on the
control footswitch allows the practitioner to work in
a perfectly sterile environment, avoiding the risk of
cross-contamination.
Once the various settings have been adjusted, the
user has no further need to touch the keypad.
Depending on the mode, the control footswitch
buttons have different functions.
I-S
URGE
mode
:
Footswitch button definitions (Fig. 11):
- 1: I-S
URGE
L
ED
motor control (ON/OFF or
progressive).
- 2: Rotation direction.
- 3: Change program (P1 to P4).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
P
IEZOTOME
mode
:
Footswitch button definition (Fig. 11):
- 1: Ultrasonics control (ON/OFF or progressive).
- 2: Flush/prime.
- 3: Change program (D1 to D4).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
N
EWTRON
mode
:
Footswitch button definition (Fig. 11):
- 1: Ultrasonics control (ON/OFF or progressive).
- 2: Flush/prime.
- 3: Change program (Soft to Boost).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
f) Technical characteristics
Manufacturer
: SATELEC
Device name
: I
MPLANT
C
ENTER
2
Electrical power supply
:
- Voltage: 100 VAC to 230 VAC.
- Frequency: 50 Hz/60 Hz.
- Rated power: 250 VA at 230 VAC.
I-S
URGE
function
Operation
:
Intermittent service: 20 sec. ON / 30 sec. OFF at 2
N.cm
Output characteristics
:
I-S
URGE
L
ED
micro-motor speed: 100 rpm to
40 000 rpm
I-S
URGE
L
ED
micro-motor torque: max. 6 N.cm
Irrigation flow rate: 10 to 120 ml/min. (nominal
value).
Adjustment in 10 ml/min. steps
Flush/prime flow rate: 120 ml/min.
P
IEZOTOME
function
Operation
:
Intermittent service: 10 min. ON / 5 min OFF
Output characteristics
:
No-load voltage: 250 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 120 ml/min. (nominal
value)
Adjustment in 10 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
N
EWTRON
function
Operation
:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics
:
No-load voltage: 150 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 40 ml/min. (nominal value)
Adjustment in 1 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
Protection
Electrical equipment class: Class 1
Electrical safety class:
Type BF in I-S
URGE
mode
Type BF in P
IEZOTOME
mode
Type BF in N
EWTRON
mode
9
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...