27
Note 1
: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
: These specifications may not be applicable in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and persons.
(a)
: The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile
telephones (cellular/cordless), mobile radios, amateur radio, AM/FM radio broadcasts and TV broadcasts
cannot be determined exactly by theory. To assess the electromagnetic environment due to fixed
radiofrequency emitters, an electromagnetic environment measurement must be made. If the measured
radiofrequency field strength in the immediate environment where the product is used exceeds the
compliance level specified above, the performance of the product must be tested to verify whether it
conforms to the specification. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the product.
(b)
: In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.
13. 4 RECOMMENDED SEPARATION DISTANCES
I
MPLANT
C
ENTER
2 is intended for use in an electromagnetic environment in which radiated radiofrequency
disturbances are controlled.
The I
MPLANT
C
ENTER
2 user and/or installer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile radiofrequency communications equipment emitters and the
I
MPLANT
C
ENTER
2, according to the maximum output power of the equipment, as recommended in the table
below.
Note 1
: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
: These specifications may not be applicable in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and persons.
Rated max. power
of the emitter (W)
Separation distance in metres (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2,5 GHz
d = 2.3
0,01
0.12 m
0.12 m
0.23 m
0,1
0.38 m
0.38 m
0.73 m
1
1.2 m
1.2 m
2.3 m
10
3.8 m
3.8 m
7.3 m
100
12 m
12 m
23 m
For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can be estimated using the
equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W) according the
manufacturer.
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...