Press button
(Fig. 3-1) or the footswitch button
(Fig. 11-2) (P
IEZOTOME
and N
EWTRON
modes) to switch
on the flush / prime function.
The flush / prime function stays on as long as the
button is pressed.
The flush can be switched on while the device is
being used.
b) Flow rate adjustment
The irrigation flow rate is adjusted using buttons
and
(Fig. 3-4 and 5).
Each time the corresponding button is pressed, the
flow rate is adjusted.
The adjusted value is displayed on the LCD screen
(Fig. 3-2) and stored in the current program.
The flow rate can be adjusted while the device is
being used.
c) Irrigation ON/OFF
The irrigation is switched ON or OFF by pressing
button
(Fig. 3-3) or the footswitch button (Fig.
11-4).
When irrigation is selected, the symbol
is
highlighted.
When irrigation is not selected, the symbol
is darkened.
7. 4 LIGHT FUNCTION
The light function is displayed on the LCD screen.
Pressing button
(Fig. 3-11) switches the light
function ON or OFF.
The light function has a default light-off time-out of
nine seconds.
Switching off the light function becomes effective
when the time-out has elapsed.
If necessary, you can lengthen the time-out (refer to
7.7).
7. 5 I-SURGE FUNCTION
a) Selecting the rotation direction
The rotation direction of the I-S
URGE
L
ED
micro-motor
is controlled by the control footswitch button (Fig.
11-2).
When clockwise rotation is selected, the symbol
is displayed on the LCD screen (Fig. 6-23).
When counterclockwise rotation is selected, an
audible signal is emitted and the symbol
is
displayed on the LCD screen (Fig. 6-23).
b) Program setting adjustment
Important:
When programming the maximum instrument end
speed, the entered value cannot exceed the pre-
programmed value.
Do not go below the minimum motor speed value
(100 rpm, 1:1 contra-angle).
Do not exceed the maximum pre-programmed
value of the instrument end torque, and do not go
below 10% of this value.
When setting the torque, please refer to the
contra-angle and implant system manufacturers’
recommendations.
Select the I-S
URGE
function by pressing button
(Fig. 3-12).
On the LCD screen, the I-S
URGE
tab must be
highlighted.
Select the program for which settings are to be
adjusted using buttons
,
,
or
(Fig. 6-
24) or using the footswitch button (Fig.11-3); the
various programmed parameter settings are
displayed.
The contra-angle
(Fig. 6-20) value is displayed
(Fig. 6-17).
The speed
(Fig. 6-21) value is displayed (Fig. 6-18).
12
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...