Configuring the Input Signal and Measurement Mode
R&S
®
EVSG1000
40
User Manual 1178.6227.02 ─ 04
Remote command:
RX 2 Source
Configures the source for RX 2. (Only available with option R&S
EVSG-B1.)
"RX 2 In"
Through the RX 2 In input on the front of the instrument, the
R&S
EVSG1000 is connected with a second receiving antenna (max.
+13 dBm) corresponding to the frequency range.
"LF In"
Through the LF In input (BNC plug) on the rear side of the instrument,
a baseband signal can be supplied for further analysis of typical AF
parameters.
Remote command:
Correction RX 1 In
Configures the correction of the RX 1 input signal (in dB).
This value is subtracted from all levels and compensates effects of upstream lines and
attenuators.
Correction RX 2 In
Configures the correction of the RX 2 input signal (in dB).
(Only available with option R&S
EVSG-B1.)
This value is subtracted from all levels and compensates effects of upstream lines and
attenuators.
4.2
Receiver Board
If a second receiver board is installed (requires the R&S
EVSG-B1 option), each
receiver board can be configured to perform different measurement tasks.
To select the receiver board to configure
1. Press the [Mode] key on the front panel.
2. Select the "RX Unit 1" or "RX Unit 2" softkey.
3. Press [Enter] or the rotary knob to confirm the selection.
To select the receiver board via VNC
1. Press the [m] key on the remote keyboard.
2. Press the [PAGE UP] or [PAGE DOWN] keys to toggle between the "RX Unit 1"
and "RX Unit 2".
3. Press [Enter] to confirm the selection.
Receiver Board