Remote Commands
R&S
®
EVSG1000
264
User Manual 1178.6227.02 ─ 04
............................................................................................... 266
.................................................................................... 267
...................................................................................... 267
MODE_FSCAN
This command switches the instrument to the "RF Spectrum" mode.
Return values:
<State>
READY.
The command was executed successfully.
FSCAN_ATTMODE?
FSCAN_ATTMODE
<Attenuation mode>
This command gets or selects the attenuation mode in the "RF Spectrum" mode.
Parameters:
<Attenuation mode>
COUPLED
Coupled attenuation
LN
Low noise attenuation
NORM
Normal attenuation (0 dB attenuation)
LD
Low distance attenuation
Return values:
<State>
READY.
The command was executed successfully.
Manual operation:
See
FSCAN_FREQCENTER?
FSCAN_FREQCENTER
<Frequency>
This command gets or sets the center frequency in "RF Spectrum" mode.
Parameters:
<Frequency>
FSCAN CENTER frequency
Default unit: kHz
Return values:
<State>
READY.
The command was executed successfully.
Manual operation:
See
RF Spectrum Mode