![Rohde & Schwarz EVSG1000 User Manual Download Page 128](http://html.mh-extra.com/html/rohde-and-schwarz/evsg1000/evsg1000_user-manual_1477821128.webp)
GBAS/SCAT-I Mode (Options R&S
EVSG-K4/-K5)
R&S
®
EVSG1000
128
User Manual 1178.6227.02 ─ 04
Results for a single frame or slot
The currently selected frame and slot are indicated by a dotted rectangle in the table.
To view the details for a particular frame or slot, turn the rotary knob to scroll through
the slots and frames. When the required slot or frame is selected, select the view you
want to see the slot results in.
For more details, see:
●
●
●
Chapter 7.3.4, "Constellation View"
●
For each slot in each displayed frame, the following information is provided in the cor-
responding table cell (for a valid GBAS burst, if available).
Burst Level Average (Av [dBm])
.................................................................................. 128
........................................................................................128
GBAS ID
4-character identifier of the ground station broadcasting the message. If there is more
than one message inside the burst, the first GBAS ID is displayed.
Burst Level Average (Av [dBm])
Arithmetic average of the burst power level measured over the period of the synchroni-
zation and ambiguity resolution field of the burst.
If no burst was detected, the arithmetic average of the noise power is displayed.
Remote command:
Application Data (App. Dat.)
Detected message types (MT) in the burst
Example: "MT:1,2": messages of type 1 and 2 detected
Status
The burst status is indicated by the color of the slot cell:
Green
The status is "OK". The received burst passed the cyclic redundancy
check (CRC).
Red
An error occured. The received burst did not pass the cyclic redun-
dancy check (CRC).
Gray
No burst received.
GBAS / SCAT-I Measurement Views and Results