Common Instrument Settings
R&S
®
EVSG1000
184
User Manual 1178.6227.02 ─ 04
The hardware status overview provides information on the current operating status of
the individual hardware components in the R&S
EVSG1000 (individually for RX1 and
RX2 boards), such as test voltages and temperatures.
The test voltages are measured continuously, internally, and checked against defined
limit values. If one of the test voltages exceeds the tolerance for more than 10
s, the
following measures are taken by the R&S
EVSG1000:
●
An error message is created in the error log (see
●
The measured test voltage is indicated in red.
●
"UNCAL" is displayed in the measurement window to indicate the measured values
may be invalid.
Remote command:
Installed Options
Access
: [Setup] > "Signal In"
Access VNC
: [n] > [F8] > [F4]
Displays a table with the currently installed option keys.
For all purchased R&S
EVSG1000 options, a license key is provided by
Rohde
&
Schwarz.
Instrument Configuration and Operating Status Settings