Graphic Measurement Modes
R&S
®
EVSG1000
115
User Manual 1178.6227.02 ─ 04
6.4
AF Time Domain Mode (Option R&S
EVSG-K12)
Access
: [Mode] > "AF Time Domain"
Access VNC
: [m] > Down arrow key
The AF Time Domain mode displays the modulation vs. time diagram of the demodula-
ted RF signal, similar to an oscilloscope. For baseband input from the "LF IN" connec-
tor, the power vs. time diagram is displayed. This mode requires the R&S
EVSG-K12
option.
The x-axis displays the measurement time. The y-axis displays the modulation depth in
percent (RF input) or the measured voltage (baseband input).
Marker functions allow you to analyze specific details in the spectrum.
Figure 6-4: AF Time Domain Mode of an ILS signal with DDM = 0,0 %
Generally, each mode is configured individually. When you switch modes, the most
recently defined settings for that mode are applied.
However, graphical results can also be displayed directly from the measurement
modes with numeric results. In this case, the settings for the current mode are applied
to the graphical results.
The following settings are displayed in the measurement settings area of the RF spec-
trum:
●
RX board ("RX1" | "RX2")
AF Time Domain Mode (Option R&S
EVSG-K12)