Graphic Measurement Modes
R&S
®
EVSG1000
118
User Manual 1178.6227.02 ─ 04
6.4.3
Configuring the Measurement Time
Access
: [MTime] > "Time/Div"
Access VNC
: [d] > "F1"
Time/Div
Defines the time range displayed in each division of the x-axis. The full x-axis range,
that is: the total measurement time, is eight times this value.
Remote command:
6.4.4
Configuring the Trigger
Access
: [Trigger]
Access VNC
: [k]
The trigger determines how often and when a new measurement is performed.
Performing a single sweep
To perform a single sweep, press [Single] (VNC: [L]).
Trigger mode
Determines how the trigger is evaluated.
"Norm"
Each time a trigger event occurs, a measurement is performed.
"Auto"
Each time a trigger event occurs, a measurement is performed.
If no trigger event occurs during a predefined waiting period, a mea-
surement is performed without a trigger. This helps you determine the
required trigger level.
"Single"
When the trigger event occurs, a single measurement is performed.
Trigger Source
Determines how often and when a new sweep is performed.
"Ext."
A measurement is performed each time a trigger signal from a con-
nected external trigger device is received. The trigger device must be
connected to the "Trigger In" connector on the rear panel of the
R&S
EVSG1000.
"Level"
A measurement is performed when the input signal reaches the
defined trigger level.
AF Time Domain Mode (Option R&S
EVSG-K12)