Common Instrument Settings
R&S
®
EVSG1000
181
User Manual 1178.6227.02 ─ 04
Example:
12.08.23
Energy Saver
Switches the energy saver setting ON and OFF. If the energy saver setting is activated,
the display is switched off when the specified time elapses. Data is still connected, only
the backlight is turned off.
To turn the backlight back on, press [Enter].
The timer can be set from one minute to 60 minutes.
Remote command:
Boot on Power Up
Automatically switches on the R&S
EVSG1000 when it is connected to an external
power supply (if not in operation already).
Remote command:
Flash size / Flash used
Indicates the size of the flash memory in the R&S
EVSG1000, and the percentage cur-
rently in use.
11.6
Instrument Configuration and Operating Status Set-
tings
The following settings concern the instrument operating status, and hardware and soft-
ware configuration.
............................................................................. 183
Power supply and battery charge state
.......................................................................186
Auto-Calibration (Cal)
Access
: [Setup] > "Cal"
Access VNC
: [n] > [F8] > [F1]
Auto-calibration corrects changes in the attenuation characteristics of the internal RF
signal paths (on both receiver boards, if available) using an internal calibration signal.
These changes can occur due to the aging process or temperature drifts. The calibra-
tion optimizes the accuracy of the level measurements.
Instrument Configuration and Operating Status Settings