Common Instrument Settings
R&S
®
EVSG1000
182
User Manual 1178.6227.02 ─ 04
The "Setup - CAL" display shows the status for each calibration step for the most
recent calibration. The first line of the result table indicates the date and time of the
auto-calibration, and the temperature during the auto-calibration.
Note:
Before starting the auto-calibration, the instrument should be in operation for at
least 15 minutes to ensure the instrument has reached its operating temperature (for
details, refer to the data sheet).
During the auto-calibration procedure, do not apply any input signal to the
R&S
EVSG1000.
To start the auto-calibration, select "Start Autocal " and press the [Enter] key.
The auto-calibration procedure is ready after a few minutes and in the result table, all
entries must have the status "OK".
Figure 11-1: Successful auto-calibration
If an error occurs during the auto-calibration process, switch the instrument off and on
again and repeat the auto-calibration process. If the error still exists, contact the
Rohde
&
Schwarz service.
It is recommended that you perfom an auto-calibration every 2 months or if the differ-
ence of the environment temperature changes by more than 10
°C.
Remote command:
Error Log
Access
: [Setup] > "Error Log"