Graphic Measurement Modes
R&S
®
EVSG1000
119
User Manual 1178.6227.02 ─ 04
Trigger Edge
For external triggering, this setting defines whether a measurement is performed when
the positive (rising) or negative (falling) edge of the trigger signal is received. Select the
key repeatedly to toggle between the two settings.
Remote command:
Trigger Level
For the
= "Level", this value specifies the level in percent the input sig-
nal must reach to trigger a measurement.
Remote command:
6.4.5
Configuring the Trace
Access
: [Meas]
Access VNC
: [a]
Trace settings determine how the display values are calculated from the measured val-
ues.
Trace mode
Defines the update mode for subsequent traces.
"Clear/Write"
Overwrite mode (default): the trace is overwritten by each sweep
"Average"
The average is determined over several sweeps
The
determines the number of averaging procedures.
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
EVSG1000 saves each trace point in the trace
memory only if the new value is greater than the previous one.
The
determines the number of sweeps to evaluate.
"RMS"
The RMS value for each trace point over several sweeps is deter-
mined and displayed.
The
determines the number of sweeps to evaluate.
Remote command:
Average Count
Determines the number of sweeps over which trace evaluation is performed (e.g. aver-
aging or maxhold).
AF Time Domain Mode (Option R&S
EVSG-K12)