How to clean dental unit surfaces
When
Part
Cleaning agent
Additional cleaning method
Dish-
washer
(65°C)
Washer-
disinfector
(93°C)
Autoclave
(134°C)
After the working
day
Upholstery
Planmeca
approved
upholstery
disinfectant
On Flexy-holder:
Suction tube
holders (incl.
rollers),
instrument holder
and
supplementary
holders
Planmeca
approved surface
disinfectant
X
On Flexy-holder:
Holder for
intraoral scanner
Planmeca
approved surface
disinfectant
Parts that should be cleaned weekly or monthly
When
Part
Cleaning method
Weekly
Upholstery
Treatment with Dürr FD 360.
Wipe away any excess oil after treatment.
Monthly
Instrument flushing
holder
Washer-disinfector (93°C) or autoclave (134°C).
See also section "Instrument flushing holder" on page
Suction tube
cleaning holder
Washer-disinfector (93°C).
See also section "Suction tube cleaning holder and
Coarse filters
Replace them.
See also section "Coarse filters" on page 242.
Instrument console
Before cleaning the instrument console, lock the touch display by pressing
the Lock button in the
Maintenance window.
When the touch display is locked, the Lock button is blue. Press the button
for 1 second to unlock the touch display. A progress bar is displayed while
you press the button.
Alternatively, the dental unit can be configured so that you can lock/unlock
the touch display by pressing the Flexy button. Contact your Planmeca
dealer.