How to clean dental unit surfaces
When
Part
Cleaning agent
Additional cleaning method
Dish-
washer
(65°C)
Washer-
disinfector
(93°C)
Autoclave
(134°C)
Balanced
instrument arms
Planmeca
approved surface
disinfectant
Hanging-tube
instrument
holders
Planmeca
approved surface
disinfectant
X
X
Control panel
Planmeca
approved surface
disinfectant
Cup fill tube
Planmeca
approved surface
disinfectant
Bowl surfaces
Mild soap- and
water solution
Planmeca
approved surface
disinfectant
X
Bowl filter
Mild soap- and
water solution
Cuspidor
Planmeca
approved surface
disinfectant
Tray tables
Planmeca
approved surface
disinfectant
Flexy-holder,
tablet holder and
suction arm
Planmeca
approved surface
disinfectant
Suction
handpieces
Planmeca
approved surface
disinfectant
X
X
Suction tubes
Planmeca
approved surface
disinfectant
Monitor
Planmeca
approved surface
disinfectant
Upholstery
Mild soap- and
water solution
29 Cleaning and disinfection