Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 27
Drive backrest upwards, then lower chair. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the backrest upwards before lowering
the chair. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 28
Planmeca Compact WE's help message.
Connect the lifter cable.
H 29
Drive chair downwards, then lower backrest.
If patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the chair downwards before lowering
the backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 30
Close water tap. Fill container with
disinfectant up to groove and attach to unit.
H 31
Switch unit off, leave disinfectant in unit
overnight (min. 8 h).
When unit is turned back on, disinfection
procedure continues.
H31.1
Remove container, fill it up to groove with
water and attach to unit.
H31.2
Attach container filled with water.
H 32
Container flushing completed.
Remove container from unit.
H 33
Open water tap.
H 34
No instruments selected when starting
instrument flushing or waterline cleaning.
Select at least one instrument and syringe
(remove from holder, bend the instrument
arms).
H 35
Instruments are locked for safety reasons.
Select instruments, place them in flushing
holder, and press Instrument flushing button
to start instrument flushing.
H35.1
Waterline cleaning cancelled or failed.
Remove container.
Make sure water tap is open. Select
instruments, place them in flushing holder
and press Instrument flushing button to start
instrument flushing.
H 36
Flushing ready. Return instruments to
instrument console.
Also bend and release syringe's instrument
arm once.
H36.1
Instrument flushing completed, but water
flow was not detected for one or more
instruments.
Run instruments for a while with spray water
to ensure instrument flushing is sufficient.
H36.8
Low water pressure during final flushing.
Perform long instrument flushing to ensure
instruments are flushed.
H 37
Waterline cleaning has been interrupted.
H 38
Low water pressure during final rinsing.
When waterline cleaning is finished, perform
long instrument flushing to ensure
instruments are flushed.
32 Help and error messages
User's manual
Planmeca Compact i 271