29 Cleaning and disinfection
29.1 Introduction
Planmeca approved surface disinfectants, upholstery disinfectants, dental
unit water and waterline disinfectants, and suction disinfectants are listed in
the document
Planmeca approved disinfectants (30007097). The document
can be found in the
NOTE
Do not use cleaning agents in aerosol or spray form directly on any surfaces.
NOTE
All parts must be cleaned before disinfecting or autoclaving them.
NOTE
If disinfectant or cleaning solution is splashed on the surfaces of the dental
unit, remove the splashes instantly with water and mild soap to avoid stains.
29.2 Dental unit surfaces
The table below lists when and how to clean the dental unit surfaces.
How to clean dental unit surfaces
When
Part
Cleaning agent
Additional cleaning method
Dish-
washer
(65°C)
Washer-
disinfector
(93°C)
Autoclave
(134°C)
After every
patient and after
the working day
Metal parts of
headrest
Planmeca
approved surface
disinfectant
Metal parts of
armrests
Planmeca
approved surface
disinfectant
Front cover of
operating light
Planmeca
approved surface
disinfectant
Handles of
operating light
Planmeca
approved surface
disinfectant
X
X
Instrument
console
Planmeca
approved surface
disinfectant
Hygienic
membrane
Planmeca
approved surface
disinfectant
X
X
Instrument hoses Planmeca
approved surface
disinfectant
29 Cleaning and disinfection
User's manual
Planmeca Compact i 235