The picture below shows the water tap when it is open.
Once the flushing cycle is finished, the LED indicator on the cuspidor's
chair side is a steady green.
7. After the long flushing cycle, make sure that the water coming from the
instruments and cup fill line is clear.
If the water is blue, run water through all water consuming instruments
and the cup fill line until the water is clear, and increase the flushing
time.
NOTE
The duration can be extended, see section "Adjusting maintenance
After this, the unit is ready for normal use.
28.6.4 Maintenance
Change the waterline disinfectant container once a year.