RPM and torque settings for different presets
Function
Preset 1
Preset 2
Preset 3
Preset 4
Preset 5
Preset 6
Drive mode
Auto
forward
Auto
reverse
Auto stop
N/A
N/A
N/A
RPM limit
260
260
900
2 000
5 000
40 000
Torque %
30
30
50
N/A
N/A
N/A
Instrument spray
Off
Off
Off
Air 70%
Water
40%
Air 70%
Water
40%
Air 70%
Water
40%
Automatic chip blow
Off
Off
Off
Off
Off
Off
Instrument light
brightness
100
100
100
100
100
100
23.5 Morita TORX micromotor
CAUTION
The user must be familiar with endodontic procedures before using the
Morita TORX micromotor.
When coupled with the CA-10RC-ENDO 10:1 handpiece, the Morita TORX
micromotor can be used together with the Morita Root ZX mini U apex
locator to measure the root canal.
Besides the basic micromotor functions, the Morita TORX micromotor has
some additional features. These are described below.
23.5.1 Torque
Press Torque to enable the torque function. The green indicator light means
that torque is enabled.
When torque is enabled, also the drive mode Autoforward is enabled. When
the torque limit is reached, the micromotor will rotate counter-clockwise
(reverse) until the torque value has decreased to 50% of the torque limit
value and then return to clockwise rotation.
NOTE
Only the drive mode Autoforward is available for the Morita TORX
micromotor.
NOTE
The foot control can be configured so that you push the left-side knob down
to enable/disable the torque function. Contact your Planmeca dealer.
NOTE
Check the indicator light on the control panel to verify that pushing the left-
side knob down activates the correct function.
23 Operating instruments
User's manual
Planmeca Compact i 145