1. Open the cap of the suction disinfectant container.
2. Fill the container with Planmeca approved suction disinfectant up to the
maximum marking (1).
3. Close the cap of the container.
When you have filled the suction disinfectant container, it is very important
that you confirm the filling of the container so that the suction disinfectant
counter is reset.
If help message H47.1 was displayed, confirm the filling by pressing the
Suction disinfectant button on the help message.
Help message H47.2 is then displayed to inform that the suction disinfectant
container is filled.
If you filled the container before help message H47.1 was displayed, confirm
the filling according to the instructions in section "Confirming suction
disinfectant container fill" on page 200.
28.6 Waterline cleaning
28.6.1 Introduction
Waterlines are a breeding ground for biofilm. Biofilm may contain bacteria
that are harmful to the patient and to the dental team. This is why the
waterlines of the dental unit should be cleaned once a week with a
disinfectant. The solution is left in the unit overnight and the waterlines of the
unit are flushed with water the next morning. The disinfectant should not be
left in the waterlines for longer than one night.
After installation, or if the dental unit has not been used for a long time, the
waterlines of the dental unit must be treated with a disinfectant for three
consecutive cycles.
If you have several dental units with Waterline Cleaning System (WCS), you
may use the same container in all of them.
28 Flushing and cleaning programs
222 Planmeca Compact i
User's manual