32.3.17 Error messages generated by control panel
Error messages generated by control panel
ERROR
CODE
EXPLANATION
EP19.3
Control panel version mismatch.
EP19.5
Control panel software upgrade has failed or memory card is
faulty.
EP21.1
Control panel display is faulty.
EP21.2
Control panel data error.
EP21.5
Unable to read Main PCB software version.
EP21.6
Control panel software not installed or software error.
Insert software update SD-card or contact service.
EP21.7
Touch panel is not working.
32.4 Viewing help and error message history
About this task
The dental unit keeps a log of all help and error messages. It can be viewed
in the
About this unit window.