The system will first identify the instruments and then each instrument
hose is flushed with air and water for 30 seconds in its turn.
During the flushing cycle the LED indicator on the cuspidor's chair side
blinks green and the LED indicator on the cuspidor's outside blinks
white.
6. Release the instrument arms (if your dental unit is equipped with
balanced instrument arms).
7. Once the flushing cycle is finished, H 36 is displayed and the LED
indicator on the cuspidor's chair side is a steady green. Remove the
instruments from the flushing holder and return them to the instrument
console. Place the flushing holder cover back to its position.
If water flow was not detected for all water consuming instruments,
H 36.1 is displayed. Run instruments for a while with spray water to
ensure that the flushing of instruments is sufficient.
28.4 Long instrument flushing
About this task
NOTE
The duration of the long flushing cycle can be adjusted, see section
"Adjusting maintenance functions" on page 200.
NOTE
In units equipped with a water heater, the long flushing fills the waterlines
and water heater with cool water and switches the water heater off.
NOTE
You can interrupt the flushing cycle by pressing Close. After the interruption,
the dental unit can be used normally.
Steps
1. Press Maintenance to go into maintenance mode.
2. Remove the cup from the cup holder. Turn the bowl away from above
the flushing holder as shown in the figure. Turn also the cup fill tube
28 Flushing and cleaning programs
User's manual
Planmeca Compact i 215