Headrest related error messages
ERROR
CODE
EXPLANATION
E82.1
Headrest software is incompatible with MCB software.
Headrest or MCB software should be upgraded.
E82.2
Incompatible headrest software.
32.3.13 ActiveAqua related error messages
ActiveAqua related error messages
ERROR
CODE
EXPLANATION
E83.1
ActiveAqua and main PCB settings vary.
Check and set water hardness (dH) and electrolysis settings in
service modes (n.181 and n.182) or in the ActiveAqua view.
Reset the counters for prefilter replacement and annual
maintenance.
E83.2
ActiveAqua electrolysis chamber asymmetrical current error.
E83.3
ActiveAqua electrolysis chamber high current error.
E83.4
ActiveAqua electrolysis chamber low current error.
E83.5
ActiveAqua RT clock malfunction.
E83.6
ActiveAqua water flow high.
E83.7
ActiveAqua level sensors missing.
Level sensors are not connected.
E83.8
ActiveAqua electrolysis chamber very high current.
E83.9
ActiveAqua water container fill timeout.
ActiveAqua can not fill the water container. Ensure that
nothing blocks the inlet water (stuck inlet water filter or flow
reducer inside the main water block) and that the water
container does not leak.
E84.1
ActiveAqua water container fill self-test fail.
ActiveAqua can not fill the water container. Ensure that
nothing blocks the inlet water.
E84.2
ActiveAqua electrolysis chamber self-test fail.
Electrolysis chamber current not detected in ActiveAqua self-
test. Check the chamber connection.
E84.3
ActiveAqua water flow self-test fail.
Water flow not detected in ActiveAqua self-test. Check the
flow meter, circulation valve and pump.