4. Press Edit next to
Water hardness (dH) to adjust the water hardness
value.
The
Program - Water dH level window opens.
5. Adjust the water hardness level by dragging the handle to a suitable
position.
The water hardness level must be set according to information received
from your local water supplier or local authority.
The value range is 0 - 40 °dH.
6. Press Edit next to
Electrolysis strength, Normal use to adjust the
electrolysis strength in situations when instuments or the cup fill function
are used.
The
Program - Water chlorine level window opens.
7. Adjust the water chlorine level by dragging the handle to a suitable
position.
The default value is 50, but is automatically modified according to the
weekly free chlorine level measurements.
The value range is 0 - 100%.
NOTE
Adjusting this setting resets the free chlorine 7-day counter.
8. Press Edit next to
Electrolysis strength, Instrument flushing to adjust the
electrolysis strength when instruments are flushed.
The
Program - Chlorine level window opens.
9. Adjust the chlorine level by dragging the handle to a suitable position.
The value of
Electrolysis strength, Instrument flushing must be at least
that of
Electrolysis strength, Normal use. Thus, if the electrolysis strength
for
Normal use = 50, the electrolysis strength for Instrument flushing
must be 50 or higher.
The value range is 0 - 100%.
10. Press OK to close the window.
30 Planmeca ActiveAqua
User's manual
Planmeca Compact i 259