Suction holder related error messages
ERROR
CODE
EXPLANATION
E48.1
No suction holders detected.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E48.2
Suction holder related error messages.
E48.3
Suction holder related error messages.
E48.4
Suction holder related error messages.
E49.1
Suction holder PCB or its cable is faulty in branch 1.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E49.2
Suction holder PCB or its cable is faulty in branch 2.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E50
Unidentified STCS jumper configuration.
E50.1
Water management system (WMS) must be set to Bottle
water off -mode when starting suction cleaning.
E50.2
Suction disinfectant pump is not running.
32.3.7 Foot control related error messages
Foot control related error messages
ERROR
CODE
EXPLANATION
E51.1
Cable from foot control is loose or connected into wrong
connector on MCB.
E51.2
Foot control is not responding.
Ensure that the foot control cable is properly attached to the
unit. If it is, contact service.
E51.3
Foot control pedal was pressed down when dental unit was
switched on.
Restart the unit without touching the foot control.
E51.4
Foot control pedal was not in home position when dental unit
was switched on.
Restart the unit without touching the foot control.
E51.5
Foot control's chair control knob was not in home position
when dental unit was switched on.
Restart the unit without touching the foot control.
E51.6
Foot control left knob is either stuck or active at power-up or
after pedal use.
E51.7
Foot control right knob is either stuck or active at power-up or
after pedal use.
E52
Foot control is of new type, update MCB software.
MCB software must be updated.
32 Help and error messages
284 Planmeca Compact i
User's manual