5. Help message H 32 is now displayed.
It means that the pre-rinsing cycle is completed and the container should
be removed. Remove possible excess water from the container into the
sink. Close the container by attaching the lid. Place the container on a
table or in a cabinet to wait for the next use.
6. Help message H 33 is displayed, meaning that the unit is waiting for the
water to be turned on.
A long instrument flushing cycle using the water from the dental unit
waterlines is automatically started after opening the tap. The long
flushing takes 4 minutes by default in a dental unit with cup fill and bowl
rinse functions.
The duration can be extended, see section "Adjusting maintenance
28 Flushing and cleaning programs