WARNING
Even though every effort has been taken to ensure patient safety, even in
case of malfunction or misuse, always make sure that the unit is properly
flushed before taking it into use.
CAUTION
Only Planmeca Planosil, Planmeca PlanPure or Alpro Bilpron disinfectant
must be used. Planmeca does not guarantee the suitability of and is not
liable for damages caused by other disinfectants.
NOTE
If you are using the Clean Water System (CWS), clean the waterlines
according to the instructions in section "Waterline cleaning" on page 261.
NOTE
Waterline Cleaning System is not available for dental units equipped with the
Planmeca ActiveAqua water treatment system.
NOTE
The dental unit waterlines should be cleaned once a week; the Maintenance
main window displays when cleaning programs were started. If Planmeca
Romexis Clinic Management software is used, detailed information on
cleaning cycles can be retrieved. For more information, see the Planmeca
Romexis user’s manual.
NOTE
Do not let the Waterline Cleaning System run over the weekend.
NOTE
The water must flow properly through the cup fill line, otherwise the unit will
not be flushed. Make sure that the cup fill line is not totally closed. The flow
through the cup fill line is recommended to be approximately 1 dl / 5 sec.
28.6.2 Starting waterline cleaning cycle
About this task
NOTE
The waterline cleaning program can be cancelled by pressing Cancel. If the
cleaning cycle is cancelled after the disinfectant container has been attached
to the dental unit, flushing is required. Help message H 35 guides you
through the flushing.
NOTE
The pop-up windows can be minimised by pressing Minimise.
Steps
1. Press Maintenance to go into maintenance mode.
28 Flushing and cleaning programs
User's manual
Planmeca Compact i 223