Foot control related error messages
ERROR
CODE
EXPLANATION
E52.1
Incompatible wireless foot control receiver software.
E52.2
Incompatible wireless foot control software.
E52.3
Incompatible foot control software.
E53
Foot control data error (cable short circuit).
E54.0
Foot control error.
E54.1
Foot control EEPROM write failure.
E54.2
Foot control EEPROM read failure.
E54.3
Pedal too far from sensors, channel A.
E54.4
Pedal too far from sensors, channel B.
E54.5
Sensor plate is tilted too much to either side, checksum error.
E54.6
Sensor plate is too close to PCB, channel A.
E54.7
Sensor plate is too close to PCB, channel B.
E54.8
Foot control re-trigger error.
E54.9
Foot control calibration error.
E55
Wireless foot control connection lost.
32.3.8 Operating light related error messages
Operating light related error messages
ERROR
CODE
EXPLANATION
E56
Operating light erroneously connected or short circuit in push
button.
E57
Incompatible SingLED software.
E58
Operating light bulb burned out.
Replace bulb.
32.3.9 Separator system related error messages
Separator system related error messages
ERROR
CODE
EXPLANATION
E59.1
Separator reports functional error.
Refer to Metasys separator documentation.
E59.2
Separator reports functional error.
Refer to Dürr separator documentation.
E60
Separator is not responding or separator cable is faulty.
E61
Separator is flooded with too much water.
E62
Newly attached separator is of new type and cannot be
controlled by MCB.