Lets you select the water supply for your dental unit. Turn the switch
toward ‘CITY’ to select domestic water, and toward ‘BOTTLE’ to select
the clean-water bottle.
2. Water bottle pressure control
The pressure in the water bottle can be adjusted by turning the knob.
The pressure can be checked from the pressure monitor and should be
between 2.5 and 2.8 bar.
3. Water bottle
When the clean-water bottle is used (water supply -switch turned toward
‘BOTTLE’), the water bottle -switch lets you turn the bottle on or off.
When the switch is toward ‘ON’, the pressure is maintained in the bottle,
and the bottle can be used.
When the switch is toward ‘OFF’, you can detach the bottle, for example,
for cleaning. Allow a few seconds for the pressure to drop before
unscrewing the bottle.
31.2 Waterline cleaning
The waterlines of the dental unit should be cleaned once a week with a
Planmeca approved waterline disinfectant. The solution is left in the unit
overnight and the waterlines of the unit are flushed with water the next
morning. The disinfectant should not be left in the waterlines for longer than
one night.
CAUTION
Only Planmeca Planosil, Planmeca PlanPure or Alpro Bilpron disinfectant
must be used. Planmeca does not guarantee the suitability of and is not
liable for damages caused by other disinfectants.
NOTE
To avoid stains, possible disinfectant splashes should be wiped away
immediately.
31.2.1 After the working day
About this task
NOTE
You can interrupt the flushing cycle by pressing Close. After the interruption,
the dental unit can be used normally.
Steps
1. Open the dental unit door.
31 Clean water system