28 Flushing and cleaning programs
28.1 When to use cleaning programs
We recommend that you run the flushing and cleaning programs as follows:
NOTE
The duration of the short and long flushing can be adjusted, see section
"Adjusting maintenance functions" on page 200.
Recommended cleaning intervals
Icon
Cleaning program
When
Duration
Short flushing
After every patient
30 seconds / instrument
(default)
Long flushing
In the morning and after
the working day
2 minutes (default)
Suction cleaning
After the working day
6 - 8 minutes
Waterline cleaning
Once a week. We
recommend that waterline
cleaning be performed
once a week after the
working day (but not over
the weekend).
2 minutes in the e
min. 8 hours affect time +
5 minutes in the morning
28.2 Before you start
NOTE
Always feed cold water to the dental unit.
NOTE
The water in the dental unit is intended for rinsing only, not for drinking.
NOTE
The water tap must be closed when the dental unit is not in use.
NOTE
If disinfectant is splashed on the surfaces of the dental unit, remove the
splashes instantly with water and mild soap to avoid stains.
In the Maintenance mode you can perform instrument flushing as well as
cleaning of the suction tubes and the dental unit’s waterlines.
The duration of the cleaning procedure depends on the dental unit
configuration and the amount of instruments and suction tubes to be
cleaned.
When the cleaning procedure of the instruments is running, the status of the
procedure is displayed on the control panel.
28 Flushing and cleaning programs
210 Planmeca Compact i
User's manual