Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 39
Syringe misplaced or its flow not detected in
waterline cleaning.
Return syringe to console. Then, select
syringe (bend instrument arm) again and
place it carefully in flushing holder.
H 40
Too many Bien-Air MX brushless
micromotors.
A maximum of two is allowed.
H 41
Insert suction tubes to suction tube cleaning
holder, add disinfectant and press green
start button (upper button).
H 41.1 Insert suction tubes to suction tube cleaning
holder.
H 41.2 Remove suction tubes from holder to start
suction cleaning.
H 42
Suction cleaning completed. Return suction
tubes to suction holder and close suction
cleaning cover.
H 42.1 Suction cleaning completed. Return suction
tubes to suction holder.
H 43
Suction cleaning cancelled. Return suction
tubes to suction holder.
H 44
Incorrect suction tube amount detected.
Return them to suction holder, place them
back to suction tube cleaning holder and
start cleaning.
H 45
Set WMS to Domestic Water Supply -mode
and wait until WMS bottle is full. Then start
suction cleaning.
Suction cleaning can not be used when
WMS is set to Bottled Water -mode.
H 46
Suction use is forbidden or there are no
suction tubes in dental unit configuration.
The value of service mode n.83 is 0.
H 47
Suction disinfectant pressure is low. (In
automatised suction cleaning: Fill suction
disinfectant container).
H 47.1 Suction disinfectant level low. Fill suction
disinfectant container.
H 47.2 Suction disinfectant container filled.
H 48
WMS must be set to Bottled water off and
Bottle on -modes before starting suction
cleaning.
Bottled water off = domestic water is used
Bottle on = bottle is used; domestic water is
run through the bottle
H 50
No cup in cup holder. Sensor does not
recognise cup when pressing cup fill button.
Place a cup on the cup holder. Or change
the setting of service mode n 115.
H 51
Bowl is above patient chair when driving
chair up.
If the unit has a bowl, move the bowl to the
rest position. If the unit has an OP delivery
arm safety switch, check that the OP
delivery arm is not blocking the chair
movement.
32 Help and error messages
272 Planmeca Compact i
User's manual