3. Press Instrument speed/power limit.
When you are programming the turbine, press Turbine speed/power
limit.
A pop-up window opens.
4. For W&H and Minetto brushless micromotors only: Press %/rpm to
select whether you want to adjust the instrument speed/power limit in
percentage points or in rpm. The appearance of the button in the
instrument view changes accordingly.
5. Adjust the instrument speed/power limit.
The minimum speed/power limit is 5 or 10% (depending on the
instrument) and the maximum is 95%. The adjustment step is 5.
If the preset value is, for example, 50%, the foot control pedal movement
controls the speed/power between 5 - 50% instead of the normal range
of 5 - 100%.
6. Press OK.
The pop-up window closes.
25.4.2 Instrument spray
Steps
1. Activate the instrument.
2. Press Program.
3. Press Instrument spray.
A pop-up window opens.
4. Adjust the flow rates for water and air.
Scaler: Adjust the flow rates for water 1 and water 2.
The minimum value of all parameters is 0 (no flow) and maximum 100.
The adjustment step is 1 in the value range 0 - 35% and 5 in the value
range 35 - 100%.
5. Press OK.
The pop-up window closes and the new values are displayed on the
control panel.
25 Programming
User's manual
Planmeca Compact i 181