CAUTION
Endodontic treatment must only be performed using the Morita TORX
micromotor's a1 - a3 presets and the Morita CA-10RC-ENDO 10:1
handpiece.
CAUTION
Always check the preset settings before using the micromotor.
The table below lists the factory presets for the Morita TORX micromotor.
These presets can be modified, see section "Modifying preset values" on
page 188.
The tolerance for RPM and torque values is +/- 5%.
NOTE
The torque limit value depends on the handpiece type.
Factory presets for Morita TORX micromotor
Setting
Preset 1 - 3
Preset a1
Preset a2
Preset a3
Preset t1 - t3
Instrument
spray
Spray 1:
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Disabled
Disabled
Disabled
Spray 1:
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Sterile water
Available
N/A
N/A
N/A
Available
Cooling air
Enabled
Disabled
Disabled
Disabled
Enabled
Automatic chip
blow
On/Off
Water: 0
Air: 60
Default: Chip
blow on
Disabled
Disabled
Disabled
On/Off
Water: 0
Air: 60
Default: Chip
blow on
Manual chip
blow
Can be
configured to
be used from
foot control
Default: On
N/A
N/A
N/A
Can be
configured to
be used from
foot control
Default: On
Rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
23 Operating instruments
148 Planmeca Compact i
User's manual
Summary of Contents for Compact i5
Page 1: ...PlanmecaCompact i5 user s manual EN 30011782...
Page 10: ...Table of contents Planmeca Compact i User s manual...
Page 325: ......