Cypher SPM
User Guide
11. Adjust view system sliders.
l
Adjust the view system slider controls to their half-
way positions.
This puts them safely out of way of the top cover.
12. Remove enclosure top.
l
Lift upward while wiggling to remove the enclosure
top. Set aside.
Do not pull forward. The SPM instrument is not
secured inside the enclosure and if it moves you may
disconnect circuit board connections.
13. Leave the enclosure as it is; we will return to it later in
Packaging the enclosure on page 233
31.7 Backpack removal and packaging
Call or otherwise contact Asylum Research and talk to us before moving anything. Chances are
that with a few phone calls and the help of this chapter you can do the job yourself. Not calling us
will likely result in instrument damage, costly repairs and downtime, and the voiding of your war-
ranty.
Materials
Part# 991.018: Cardboard Carton 18" x 18" x 6"
Part# 990.038: Foam (backpack)
Tools
Part# 290.134: 3/32” Hex driver
Final weight
15 lbs / 7 kg
Minimum preparation System Shut Down (see
Complete system shutdown on page 218
).
Enclosure Top Removal (see
Removing the enclosure top cover on
).
Page 225 of 243
31 Shipping or moving