Cypher SPM
User Guide
9. Secure the cell body.
l
Locate three 2-56 X 1/8” screws and the 1/16” hex
driver tool.
l
Fasten the screws as shown, using only your fin-
gertips to hold the tool. This will prevent over tight-
ening.
10. As shown, insert the heater cable connector and push it flush to the rim.
11. Check the installation.
l
Your scanner should now look like the photo to the
right.
The sample stage top surface sits quite deep as shown,
below the glass sidewalls of the cell body. View the
stage through the windows on the sides of the cell and
make sure you cannot see it.
If the sample stage is not seated properly, you will
crush it when you go to insert the cantilever holder.
10.9.5 Replacing the scanner
1. Refer to
Replacing the scanner on page 111
.
2. When pressing the scanner against the back of the chassis, you should sense a hard stop of
metal touching against metal.
This requires a little extra pressure to compress some springs behind the copper plate at the
back of the scanner.
If you do not press the scanner all the way into the chassis, thermal performance might suf-
fer when using cooling or heating stages.
10 Cell Body and Sample Stage Guide
Page 90 of 243