Cypher SPM
User Guide
5. As shown, guide the connector under the ring and leave it sitting loose on the top of the scan-
ner.
6. Attach gas lines.
l
Using only your fingers to tighten the screws, attach
the gas lines to the cell body as shown.
l
Tighten until snug, don't overdo it.
7. Seat the sample stage.
A properly seated sample stage is centered in the cell body.
l
Lower the dovetail connection at the bottom of the
stage into the receiving hole on top of the scanner.
l
Press lightly onto the top of the sample stage until
the sample stage sinks down into the hole.
l
Check that the cell body is seated on the scanner
engage ring.
8. Tighten sample stage.
l
Press down on the sample stage to keep it flush
against the scanner.
l
Tighten the screw using the 5/64” driver. This will
take about three turns before you feel resistance.
l
Tighten snug, using only your fingertips to handle
the tool.
Page 89 of 243
10 Cell Body and Sample Stage Guide