3. Install the applications board.
l
Place the applications board over the connections on
the top right side of the Environmental scanner.
4. Seat the board.
l
Gently push the board onto the connectors.
5. Fit the cell body into the scanner and secure it to the engage stage ring with 3, 2-56 x 1/8”
screws.
6. Secure the stage base into the scanner.
Note:
The humidity cell is delivered with an Ambient Stage already installed. The humidity
cell body is compatible with all the Environmental Scanner Stages. If your experiment
requires temperature control then the ambient stage can be substituted for one of the act-
ive stages.
7. Connect the pigtail cable from the humidity cell body to the applications board.
l
Gently push the plug on the cable into the connector
on the board.