Cypher SPM
User Guide
6. Release the cantilever holder.
l
Locate the tool with yellow tip in the chassis to the
left of the scanner.
l
Use tool to loosen the two screws clamping the can-
tilever holder. One turn counterclockwise should be
enough - do not completely remove the screws.
l
Replace the tool.
7. Rotate the cantilever holder.
l
Place your fingers on the two side cantilever holder
handles.
l
Rotate counterclockwise a few degrees.
l
The cantilever holder circuit board comes out of its
mating connector (1). Stop when the board has
cleared the connector.
l
Notice the screws come away from the cutouts in the
cantilever holder (2).
8. Remove the cantilever holder.
l
After the board has cleared the connector, carefully
wiggle the cantilever holder up and out.
l
The resistance you feel as you remove the cantilever
holder comes is the cantilever holder O-ring sliding
out from the cell body.
Tip:
After the O-ring clears the cell body, be ready
for the resistance to drop suddenly. Keep a firm
grip on both handles to keep from flinging the
cantilever holder across the room!
9. Set aside the cantilever holder.
l
Set the cantilever holder on its handles with the can-
tilever facing up so that you do not crush the can-
tilever.
21 Removing the cantilever holder on an environmental scanner
Page 153 of 243