Cypher SPM
User Guide
10.8.2 Attach cell body and sample stage
1. Prepare parts.
l
Sample stage on the left.
l
Cell body in the middle
l
Membrane clamping ring and screws to the right.
More screws (00-90 X 1/8”) can be found in the Envir-
onmental Scanner accessory kit, see
2. Place the stage.
l
Note the three small holes in the “petals” on the dia-
phragm perimeter.
l
Align those with the three pins on the bottom of the
cell. The parts can only go together one way.
l
Assist the diaphragm so the pins go through the
holes.
3. Place the ring.
l
As shown in the photos, place the ring.
l
The ring has three small holes that line up with the
pins.
l
Seat the ring flush against the diaphragm.
Tip:
The ring has a smooth side and a side with
raised metal features. When the ring is properly
placed, the smooth side is showing and the raised
features face the diaphragm.
4. Place the screws as shown, ensuring that the ring is the correct way up.
10 Cell Body and Sample Stage Guide