Cypher SPM
User Guide
18
18
Removing The Cantilever Holder On A Standard
Scanner
Note:
If you own the environmental scanner, please follow the tutorial in
tilever holder on an environmental scanner on page 151
.
1. Raise the cantilever holder.
l
Rotate the Engage Control knob on the Cypher clock-
wise and hold it until the cantilever holder is far from
the sample or is at its upper limit of travel.
If you turn the knob counterclockwise you will
lower the cantilever holder, risking collision with the
sample and damage to the scanner.
l
For safety reasons we recommend making motor
moves with the door closed. Beware of pinch points.
2. Lift the door latch and open the enclosure door.
18 Removing the cantilever holder on a standard scanner