Cypher SPM
User Guide
5. Make sure the locking lever on the chassis is in the raised position.
6. Set the scanner on the rails.
l
Grasp both sides of the scanner and remove. Place
your fingers under the two dovetail rails as shown in
the photo.
l
Use care because the scanner is heavy.
l
Lift the scanner as shown and set it on the rails.
l
Push it in far enough so it sits stably on the rails.
7. Secure the scanner connector.
l
Press the connector into its mating socket.
l
Turn the knob on the right side of the scanner clock-
wise until firmly tight.
8. Push the scanner all the way in until it connects with the chassis.
9. Lock the scanner.
l
Continue to apply pressure against the scanner.
l
Lower the lever to the right of the scanner to lock it
into place.
14 Scanner Exchange