Cypher SPM
User Guide
1. Locate parts.
l
Locate the Cleaning Cup (114.820).
l
Select the cantilever holder you want to clean.
l
Optionally, remove the screw and clip (see
).
2. Attach holder to cup.
l
Make sure the screws are loose enough.
l
With fingers on both handles of the cantilever
holder, use firm pressure to wiggle the holder down
into the cup.
l
Rotate the cantilever holder clockwise until you feel a
hard stop.
l
Tighten the screws lightly.
Note:
A perfusion cantilever holder should have
plugged fluid ports before proceeding to the next
step.
3. Add cleaning fluid.
l
Pour a small amount of solvent or water into the
cup. We recommend ethyl or isopropyl alcohol or
deionized water.
Note:
Acetone and Methylene Chloride and other
aggressive solvents should not be used since they
can attack epoxy which is exposed when the can-
tilever holder clip is removed.
4. Clean.
l
Use a soft swab to clean the surface of the cantilever
holder.
l
Rinse and repeat as desired.
5. Blow dry with compressed filtered air.
6. Remove the holder from the cup.
7. Blow dry around the large O-ring, some solvent may have become trapped in the outer peri-
meter O-ring groove.
9 Environmental scanner cantilever holders