11. After the cantilever holder is fully inserted, use the ball head wrench to gently tighten the
clamp.
12. Remove the old cantilever.
l
Position the changing station as shown, on a flat
hard surface.
l
Take the tweezers in your dominant hand.
l
Press down on the station, as shown, with your
other hand. This depresses a button on the bottom
of the station which drives a pin up under the can-
tilever retaining clip.
l
Remove the cantilever and release pressure on the
station.
l
Inspect the cantilever area for tiny silicon grit and
blow clean with compressed air if necessary.
13. Select the new cantilever.
l
Select a new cantilever and pick it up with tweezers.
l
Close the box.
Ruining $1k of levers by putting your hand on an
open box is not unheard of.
l
If your lab saves some old cantilevers, consider prac-
ticing with one before using a new cantilever.
Tip:
It can be useful to first lay the chip down on a
non-sticky surface and re-grip it before continuing.