Cypher SPM
User Guide
4. Secure sample chamber to scanner with two screws.
l
Secure the sample chamber with screw number 1. Do
not over-tighten.
l
If your cantilever holder has a hole in the position
marked (2), secure that screw also.
Note:
Some older equipment does not grant
access to (2). In that case, skip this screw.
5. Remove the bar.
l
Unscrew the bar and remove it as shown.
l
Ensure the fastening screws are in place but not
screwed all the way down, as they will be used to
secure the cantilever holder in the next step.
During these steps, the cantilever holder could be
pulled loose, exposing your sample.
6. Rotate the cantilever holder into position and secure.
l
While pressing down to maintain the O-ring seal,
rotate the holder clockwise so that the circuit board
seats with its connector and the screws line up with
their grooves on the holder.
l
Tighten screws.
7. Put the final screw into place, securing the cell body to the scanner.
l
With the holder in place and fastened to the sample
chamber, tighten the third and final screw that
fastens the sample chamber to the scanner.
12 Glovebox Protocols
Page 104 of 243