7. Rotate the cantilever holder.
l
Place your fingers on the two side cantilever holder
handles.
l
Rotate counterclockwise a few degrees.
l
The cantilever holder circuit board comes out of its
mating connector (1). Stop when the board has
cleared the connector.
l
Notice the screws come away from the cutouts in the
cantilever holder (2).
8. Remove the cantilever holder.
l
After the board has cleared the connector, carefully
wiggle the cantilever holder up and out.
l
The resistance you feel as you remove the cantilever
holder comes is the cantilever holder O-ring sliding
out from the cell body.
Tip:
After the O-ring clears the cell body, be ready
for the resistance to drop suddenly. Keep a firm
grip on both handles to keep from flinging the
cantilever holder across the room!
9. Set aside the cantilever holder.
l
Set the cantilever holder on its handles with the can-
tilever facing up so that you do not crush the can-
tilever.
10. Locate your cantilever holder.
l
Identify the correct cantilever holder.
l
This demo uses the standard gas holder, part#
901.758.
Tip:
To learn more about cantilever holders for the
environmental scanner, see
ES cantilever holders on page 54