3. Remove fittings while holding manifold.
l
Pull the manifold forward a little. Pay attention not
to apply tension to the tubing or the wires.
l
Remove the fitting using fingers or the tool shown in
previous steps.
l
When the tubing has been replaced, reverse attach
the manifold again with its screws.
11.3 Manifold lab-side connection
1. Locate the manifold port.
l
The lab-side port of the valve manifold is shown in
the photo.
2. Prepare tubing.
l
Attach the tubing with fitting to the fitting tool as
shown.
l
The tubing should have a small amountt extending
beyond the ferrule. This space holds the O-ring
(230.039)