Cypher SPM
User Guide
1. Make sure the right door is still open.
2. Note alignment pins.
l
Note the pins which help the top cover align and find
its proper place.
l
When the top cover is removed, the weight of the
doors can bend the enclosure's vertical support out-
ward. When the cover is pressed in place, these pins
pull everything back into alignment.
3. Adjust view system sliders.
l
Adjust the view system slider controls to their half-
way positions.
This puts them safely out of way of the top cover.
4. Lower the rear of the top cover down to fit it into the enclosure.
5. Lower the front of the top cover down and fit it onto the two alignment pins.
6. Adjust the position of the cover until it fits on the enclosure without any force applied.
7. Close right enclosure door.
l
Close the right enclosure door.
l
While pressing it firmly closed to compress the rub-
ber seal, install the two screws at the bottom of the
door.
These screws were removed when the cover was ori-
ginally removed. See
8. Install the upper screw on the right hand door.
This screw was removed when the cover was originally removed. See
9. Install the upper screw on the left hand door.
This screw was removed when the cover was originally removed. See
10. Install the four rear screws into the top cover.
These screws were removed when the cover was originally removed. See
15 Light tubes