Cypher SPM
User Guide
16. Tighten the nuts.
l
There is one nut on each side.
l
Turn the nut back toward the screw head and
tighten it with pliers (pliers not provided with sys-
tem).
l
SCREW INFO: 10-32 x1 1/2 or1 3/4 BHCS SS, depend-
ing on which guard you have & 2 10-32 Kepnuts.
17. Once guard is on, lift the chassis with the side handles and place it into the bottom foam/box
carefully.
Note:
The chassis weighs approximately 51 lbs.
18. Place a plastic bag around the chassis once it is in the foam.
19. Place the top foam atop the chassis.
20. Place the 1/8" hex wrench L-key in the foam cut-out or tape it to the top foam.
21. Close the box and tape it shut.
The scanner box should remain upright at all times if possible when shipping on crate or pallet.
31.9 Packaging the enclosure
Call or otherwise contact Asylum Research and talk to us before moving anything. Chances are
that with a few phone calls and the help of this chapter you can do the job yourself. Not calling us
will likely result in instrument damage, costly repairs and downtime, and the voiding of your war-
ranty.
Materials
Part# 991.015: Cardboard Carton 21” X 20” X 21”
991.015: Foam (Cypher enclosure)
Kapton Tape (1/2” or 3/4”)
Tools
Part# 290.134: 3/32” hex driver
Final weight
90 lbs / 41 kg
Minimum preparation System shutdown (see
Complete system shutdown on page 218
).
Scanner removal (see
Scanner removal and packaging on page 218
).
Enclosure top removal (see
Removing the enclosure top cover on
).
Backpack removal (see
Backpack removal and packaging on page 225
).
Chassis removal (see
Chassis removal and packaging on page 227
).
Page 233 of 243
31 Shipping or moving