Single tests in PAT testing mode
76
LIMIT
DC
Maximum voltage
:
113 Vpeak,
141 Vpeak,
None.
Test circuit and measurement procedure for No load voltage measurement
For more information refer to chapter
Measurements according to IEC/ EN
60974-4
, paragraph
No load voltage
in 3-phase adapter instruction manual.
Examples of No load voltage measurement results
Displayed results:
Main result ............. no load a.c. or d.c. peak voltage
Sub-
result…………. no load r.m.s. voltage
Notes:
During the test, a mains voltage is connected to the welding machine. Consider
safety precautions.
Consider any displayed warning before starting measurement!
6.3.10 Clamp current test
Description in chapter
6.2.9 Clamp current test
can be used as reference.
6.3.11 Functional test
Descriptions in chapter
6.2.11 Power/ Functional tests
can be used as reference.