MI 3321 MultiservicerXA
Appendix A
– Preprogrammed autotests
183
Pre-
programmed autotest sequences table (cont’d)
Autotest shortcut
code
05
06
07
08
Cl_2_Iso
Cl_2_Ibs
Cl_1_IsoIa
Cl1_IsoIaBLT
Visual test
Earth bond
Output
200 mA
200 mA
Limit
0.30
0.30
Time
5 s
5 s
Insulation
Output
500 V
500 V
Limit
1.00 M
1.00 M
Time
5 s
5 s
Insulation
(probe)
Output
500 V
500 V
Limit
2.00 M
2.00 M
Time
5 s
5 s
Sub leakage
Output
Limit
Time
Sub leakage
(probe)
Output
40 V
Limit
0.50 mA
Time
5 s
Differential
leakage
Output
230 V
230 V
Limit
3.50 mA
3.50 mA
Time
180 s
180 s
Touch
leakage
Output
230 V
230 V
Limit
0.50 mA
0.50 mA
Time
180 s
180 s
Power
Output
230 V
230 V
230 V
230 V
Limit
Time
180 s
180 s
180 s
180 s
TRMS
clamp
current
Output
Limit
Time
Polarity test